Electrical Co n n ectio n s
(see Figure 2 and 8)
All wiring must conform to the National Electric Code and any applicable local codes. The Chiller must be
permanently wired by means of electrical conduit to a properly fused disconnect of proper amperage or wired to
a properly rated power cord and plugged into an outlet with the appropriate disconnect and amperage rating.
The electrical junction box, located on the back panel of the Chiller , includes a four terminal strip for power
supply connection.
The data plate, located next to the junction box, includes the actual voltage, phase, and amperage of the Chiller.
CAUTION: On three-phase applications, it is important that the rotation of the pump, when
supplied , is correct. Operating the pump in reverse for more than a few seconds will result
in permanent pump damage. When the pump is operating, the shaft rotation
must
match the
direction indicated on the pump housing. If the rotation is incorrect, reverse two of the three
incoming power leads.
CHILLER OPERATION
CONTROL PANEL
(see Figure 3)
FIGURE 3. CONTROL PANEL
CONTROL
POWER
BYPASS
SAFETY
BLUE INDICATOR
RED INDICATOR
POWER SWITCH
TEMPERATURE
CONTROLLER
PRESS GAUGE
1.
TEMPERATURE CONTROLLER
(see Figure 4)
The TEMPERATURE CONTROLLER uses a PID control algorithm to precisely monitor and control the pro-
cess set point temperature. The Unit has a large LED READOUT that displays temperature. For adjusting
set point and programming options, see section on TEMPERATURE CONTROLLER OPERATION.